Key Takeaways
- Most pressure gauge failures trace to six causes: pulsation and vibration, temperature, pressure spikes, clogging, corrosion, and poor installation or handling. Each has a known fix at the selection or installation stage.
- A gauge range of roughly twice the maximum operating pressure avoids both early fatigue and poor resolution.
- Any suspicious pointer behaviour justifies a full accuracy check. Whether to recalibrate or replace depends on how critical the gauge is and how it has failed.
- A large span shift, a large zero shift, corrosion, leakage or heat exposure usually means replace, not repair.
A pressure gauge rarely fails at random. Most failures come from a mismatch between the gauge and its service, so fixing the cause matters as much as fixing the gauge. This guide covers the causes, how to prevent them, and when to recalibrate, repair or replace. For the visible warning signs, see our companion post on seven telltale signs your pressure gauges need calibration or repair.
What Causes Pressure Gauge Failure?
Ashcroft’s guidance lists six causes, each with a standard remedy (Ashcroft, six reasons gauges fail audits).
| Cause | What it does | Typical prevention |
|---|---|---|
| Pulsation and vibration | Pointer flutter and internal wear | Liquid-filled gauge, remote mounting, dampening accessories |
| Temperature outside limits | Accuracy drift, hardened gaskets, discoloured dials | Remote mount with capillary, or a siphon |
| Pressure spikes | Dented pointers, ruptured Bourdon tubes | Internal stops, pressure-limiting valves, higher-range gauge |
| Clogging | Particulates restrict pressure transmission | Diaphragm seals, isolation rings, flushing connections |
| Corrosion | Media attacks the sensing element | Check material compatibility, or use a diaphragm seal |
| Poor installation or handling | Over-torqued connections, damaged cases | Correct wrenches, no torque on the case, routine inspection |
Pulsation deserves a special mention. High-frequency pulsation is hard to detect, and the only symptom may be an upscale zero shift from movement wear (Ashcroft, Pressure Gauge Installation, Operation and Maintenance). A gauge that keeps drifting upscale between calibrations may be telling you about pulsation, not age.
Do Corrosive, Dirty or Oxidizing Media Cause Failures?
They do, and the fix is usually at selection. Ashcroft recommends a diaphragm seal for media that is corrosive to the sensing element, carries heavy particulates such as slurries, or is very viscous, including media that harden at room temperature. The fill matters too: glycerin suits most applications within a usable range of 20 to 180°F (about -7 to 82°C), silicone widens it to -40 to 250°F (about -40 to 121°C), and oxidizing media need a halocarbon fill and specially cleaned gauges (Ashcroft manual, above). A gauge with the wrong fill or wetted material will fail, however well it was calibrated.
How Does Temperature Affect a Pressure Gauge?
Heat and cold change the reading. Ashcroft’s rule of thumb for dry gauges is a change of 0.5% of full scale for every 40°F (about 22°C) away from 75°F (about 24°C), and double that allowance for liquid-filled or sealed cases, with very significant errors above 250°F (about 121°C) (Ashcroft manual, above). For steam, a water-filled siphon between gauge and line keeps live steam out of the Bourdon tube, and a diaphragm seal is the alternative if condensate could freeze. Heat matters in UAE summers too, because outdoor gauges in direct sun can run hotter than the surrounding air.
Is Your Gauge Range Part of the Problem?
Often, yes. Ashcroft advises a range of approximately twice the maximum operating pressure. Too low a range shortens the fatigue life of the elastic element and exposes the gauge to overpressure damage from transients. Too high a range gives poor resolution (Ashcroft manual, above). If a gauge has failed repeatedly, check its range against the real operating pressure before buying another identical one.
When Should You Recalibrate, Repair or Replace?
Any suspicious pointer behaviour warrants a full accuracy check, and the same manual notes that a user may also set a routine bench-inspection frequency. A check takes readings at about five equally spaced points, upscale and downscale, before and after lightly tapping the gauge to remove friction. Whether to bring an out-of-tolerance gauge back depends on how critical it is. Ashcroft’s example: a commercial gauge only 0.5% beyond specification may not be worth the time and cost, while a 0.25% test gauge found 0.1% out should be recalibrated. Our guide to what a pressure gauge calibration service covers explains the process and how often it’s needed.
Ashcroft’s replacement criteria are specific: a span shift above 10% (possible wall thinning from corrosion), a zero shift above 25% (likely significant overpressure), more than a million pressure cycles with significant pointer movement, any corrosion or leakage, exposure to high temperature, worn movement, damaged threads, and loss of fill in liquid-filled gauges. A smaller zero error after a drop or overpressure can often be corrected by repositioning the pointer, but the cause still needs fixing.
What Should You Do First With a Failing Gauge?
Don’t trust its reading. Cross-check against another gauge on the same line if there is one, and follow your site’s isolation procedure before removing it. Check the pointer at zero with no pressure applied, then look for fill leakage, case damage and signs of media leaking through the connection. Tag the gauge, note its serial number and the symptom, and have it assessed rather than returned to service.
Frequently Asked Questions
Usually because the Bourdon tube has taken a permanent set after a drop or an overpressure event. Have it checked, and find out what caused the overpressure.
Pulsation or vibration. Remedies include a liquid-filled gauge, remote mounting or a dampening accessory such as a snubber.
Sometimes. A technician assesses it first, and any repair is followed by a full calibration. Gauges with heavy corrosion, leakage or large zero or span shifts are usually better replaced.
Match the gauge to the service: correct range, fill, material and accessories for the pulsation, temperature and media involved. Replacing like with like repeats the failure.